Manufacturer intervals, original parts, official dealer technicians trained on the range.

Exact intervals depend on the model and environment: your unit's manual is the reference.
Schedule maintenance| Smart suction: draw a disinfecting, descaling Puli-Jet Plus 2.0 solution through with the Pulse Cleaner. | every evening |
| Source: Cattani documentation, manual smart 04 2024, p.19 | |
| Smart suction: place the disinfecting anti-foam tablets in the chair filters. | every day |
| Source: Cattani documentation, manual smart 04 2024, p.19 | |
| Smart suction: check the screen for any alarm signals; if a danger warning appears, contact the technician. | every day |
| Source: Cattani documentation, manual smart 04 2024, p.19 | |
| Smart suction: clean the suction unit filter. | periodically |
| Source: Cattani documentation, manual smart 04 2024, p.19 | |
| Smart suction: check that nothing obstructs the suction unit's ventilation. | occasionally |
| Source: Cattani documentation, manual smart 04 2024, p.19 | |
| Smart suction: check the maximum recorded temperatures and all alarm signals. | every 12 months |
| Source: Cattani documentation, manual smart 04 2024, p.20 | |
| Smart suction: replace the HEPA H14 absolute filter on the air outlet. | every 12 months, if routine maintenance is not enough |
| Source: Cattani documentation, xst agg poster turbo e micro smart man straord sett 2023, p.2 | |
| Smart suction: check the operation of the centrifugal separator and the recirculation valve. | every 18 to 24 months |
| Source: Cattani documentation, manual smart 04 2024, p.21 | |
| Smart suction: replace the motor bearings if necessary. | every 10,000 operating hours, if needed |
| Source: Cattani documentation, xst agg poster turbo e micro smart man straord sett 2023, p.2 | |
| Routine maintenance by trained practice staff; extraordinary maintenance by a trained technician, with original parts. | continuously |
| Source: Cattani documentation, manual smart 04 2024, p.19 | |
| Hydrocyclone: replace the dirty filter. | every day |
| Source: Cattani documentation, manual istruzione idrociclone iso 18 web, p.19 | |
| Hydrocyclone: clean and disinfect the unit with a product specific to suction systems and amalgam separators. | end of each day |
| Source: Cattani documentation, manual istruzione idrociclone iso 18 web, p.19 | |
| Hydrocyclone: remove the upper section and clean all its parts, especially the five orifices. | every year |
| Source: Cattani documentation, manual istruzione idrociclone iso 18 web, p.20 | |
| Amalgam container: plan its replacement at 95%; waiting until 100% can interrupt the practice's work. | at the 95% alarm |
| Source: Cattani UK & Ireland brochure, p.7 | |
| Aspi-Jet: clean the system with Puli-Jet Plus 2.0 before the lunch break. | every lunchtime |
| Source: Cattani documentation, manual aspi jet ed 10 2021, p.15 | |
| Aspi-Jet: clean the filters and the unit with hot water and Puli-Jet Plus 2.0. | end of each day |
| Source: Cattani documentation, manual aspi jet ed 10 2021, p.15 | |
| Aspi-Jet: clean jars, lids and probes with a sponge and Puli-Jet diluted in hot water. | every week |
| Source: Cattani documentation, manual aspi jet ed 10 2021, p.14 | |
| Compressor: open the tap under the tank; if there is moisture, call a technician. | periodically |
| Source: Cattani documentation, manual compressori 1 2 3 cil 02 2023, p.13 | |
| Compressor: check the equipment room temperature (between +5 °C and +35 °C). | regularly |
| Source: Cattani documentation, postermanutenzordin 02 2023 web, p.2 | |
| Compressor: check the current draw with a clamp meter. | every 6 months |
| Source: Cattani documentation, manual compressori 1 2 3 cil 02 2023, p.14 | |
| Compressor: check that the fans are working. | every 12 months |
| Source: Cattani documentation, manual compressori 1 2 3 cil 02 2023, p.13 | |
| Compressor: in a clean environment, clean the head filters with compressed air and replace them. | every 12 months |
| Source: Cattani documentation, postermanutenzionestraordinaria, p.2 | |
| Compressor: check the charging time (from 6 to 8 bar: 45-50 s). | every 12 months |
| Source: Cattani documentation, postermanutenzionestraordinaria, p.2 | |
| Compressor: bearings, service life of about 10,000 hours; anti-vibration mounts, about 5 years or 5,000 hours. | based on operating hours |
| Source: Cattani documentation, postermanutenzionestraordinaria, p.2 | |
| Compressor: routine maintenance to be carried out by a trained technician. | continuously |
| Source: Cattani documentation, manual compressori 1 2 3 cil 02 2023, p.13 | |
Answers taken from Cattani documentation. Another question? A specialist will answer.
Every evening, draw a Puli-Jet Plus 2.0 solution through with the Pulse Cleaner; place the anti-foam tablets in the chair filters; check the on-screen alarms.
Source: Cattani documentation, manual smart 04 2024, p.19
It reduces bacterial build-up in the system and prevents blockages and bad odors.
Source: Cattani UK & Ireland brochure, p.2
Routine maintenance is carried out by trained practice staff. Extraordinary maintenance is for a trained technician, with original parts.
Source: Cattani documentation, manual smart 04 2024, p.19
Air passes through silica gel, then a sintered bronze filter and an antibacterial HEPA H14 filter before entering the tank.
Source: Cattani documentation, manual compressori 1 2 3 cil 02 2023, p.9
Compressors: 3 years from sale. Aspi-Jet: 1 year. Blok-Jet Silent: suction 1 year, compressor 3 years. Always subject to return of the warranty coupon.
Source: Cattani documentation, manual compressori 1 2 3 cil 02 2023, p.4
Using products that are unsuitable or different from those specified by Cattani, incorrect use, or work by unauthorized persons.
Source: Cattani documentation, manual smart 04 2024, p.36
Describe your practice: a Cattani specialist replies with a costed configuration.